Team leaders at Leica Biosystems created a novel methodology to match the RGB values generated when D50 light passes through pathology tissue with RGB interpretation by monitors used in digital pathology.

Digital pathology is quickly growing across the globe as it offers increasing benefits to standardize and optimize the pathology lab workflow. Pathologists can now retrieve their cases in digital form and review and sign off cases using a PC and monitor. However, digital pathology companies offer many different monitors, depending on the company. Pathologists are often frustrated by the color variation of a digital image as it may not match the microscope color experience They often prefer a color that looks as close as possible to the typical experience using a microscope.

How can we color match the image viewed on the monitor to what is seen in the microscope?

The problem is that no company had calibrated the color of the histology stain to the color of the monitor and color of the whole slide image. Our innovation team worked together to solve this issue.

Image Capture RGB values versus Display (Monitor) RGB values

When an Aperio digital pathology scanner scans a slide it produces a whole slide image that contains calibrated red-green-blue (RGB) values because the image is a color image. When the image is viewed on a monitor, the monitor also has its own separate calibrated RGB values (for example sRGB calibration by the manufacturer) because it is a color monitor. The problem is that image capture RGB values generated during image acquisition do not match the monitor display RGB values.

Many customers simply assume image capture RGB values will match the monitor RGB values. But this is not the case. 

A novel solution to calibrate color using Histology slides (patent pending). 

The goal is to reproduce what is seen in the microscope so that the same information is presented to the pathologist. In order to do this, the scanner and monitor need to be matched properly in order to produce accurate color. Leica Biosystems innovation solved this by creating a novel method, which includes the following:

  • Input color is measured on a histology slide using a patented spectral method (standard D50 light used in microscopes).
  • Output color is measured on the monitor display. Small patches of a whole slide image are magnified on the monitor and a colorimeter/spectrophotometer is used to measure the color of light for each patch.
  • The scanner and monitor display are then calibrated separately for accurate color.
  • An ICC profile is created to convert between RGB values across scanner and monitor to produce accurate color.
  • Each monitor brand and model has its own ICC profile, and Aperio viewers calibrate image color against that ICC profile for consistent visualization.

Pathologists Benefit as Monitor Color Matches Histology Slides and Microscope

As digital pathology ramps up its adoption across the globe, it is important to consider the color of the image that is displayed on the monitor as Pathologists prefer that the color match the microscope (D50 light passing histology stained slides). By utilizing Leica Biosystems novel calibration methodology an ICC profile has been created to convert between RGB and color so that when a Pathologist views the image on a monitor it matches the color to histology stained slides.

For Research Use Only. Not for Use in Diagnostic Procedures


발표자 소개

Allen Olson
Allen Olson , PhD, Technical Lead Engineer and Scientist

Dr. Allen Olson is a Technical Lead Engineer and Scientist at Leica Biosystems and has been with the company since the early days of Aperio. His subject matter specialties include algorithms for automating whole slide scanning and color calibration of scanner and display. He holds a PhD in Earth Sciences (Geophysics) from Scripps Institution of Oceanography and a BA in Applied Mechanics and Engineering Sciences from the University of California San Diego. Dr. Olson is co-inventor on numerous patents related to whole slide imaging and color.

Lance Mikus
Lance Mikus , MBA

Lance Mikus earned his bachelor’s degree from UCLA. He worked as a scientist and lab manager at the UW-Madison Medical School, resulting in several publications, and went on to earn an MBA focused on Entrepreneurship and Marketing from the UW-Madison School of Business. Mr. Mikus has 20 years of commercial operations experience in the healthcare industry. He also holds a green belt in six sigma quality practices, is a certified practitioner in launch excellence and has been awarded 5 patents.

Related Content

Leica Biosystems 콘텐츠는 Leica Biosystems 웹사이트 이용 약관의 적용을 받으며, 이용 약관은 다음에서 확인할 수 있습니다. 법적고지. 라이카 바이오시스템즈 웨비나, 교육 프레젠테이션 및 관련 자료는 특별 주제 관련 일반 정보를 제공하지만 의료, 규정 또는 법률 상담으로 제공되지 않으며 해석되어서는 안 됩니다. 관점과 의견은 발표자/저자의 개인 관점과 의견이며 라이카 바이오시스템즈, 그 직원 또는 대행사의 관점이나 의견을 나타내거나 반영하지 않습니다. 제3자 자원 또는 콘텐츠에 대한 액세스를 제공하는 콘텐츠에 포함된 모든 링크는 오직 편의를 위해 제공됩니다.

모든 제품 사용에 다양한 제품 및 장치의 제품 정보 가이드, 부속 문서 및 작동 설명서를 참조해야 합니다.

Copyright © 2024 Leica Biosystems division of Leica Microsystems, Inc. and its Leica Biosystems affiliates. All rights reserved. LEICA and the Leica Logo are registered trademarks of Leica Microsystems IR GmbH.